Business Type:
Manufacturer/Factory
Establishment:
2012
R&D Capacity:
ODM, OEM, Other
OEM/ODM Service
Sample Available

Hematology Analyzer, Medical Equipment, Analyzer manufacturer / supplier in China, offering Super Quality Auto Hematology Analyzer, Infant CPAP Ventilator, Mother  & Fetal Monitor  and so on.

1
    YRS
General Supplier
Supplier Homepage Furniture Outdoor Furniture AJ-2202 Portable Medical Ventilator Portable Patient Ventilator Portable Ambulance Ventilator

AJ-2202 Portable Medical Ventilator Portable Patient Ventilator Portable Ambulance Ventilator

Get Latest Price
Min. Order / Reference FOB Price
1 Unit US $2000.00/ Unit
Local Area: Anqing, Anhui, China
R&D Capacity: OEM, ODM, Other
Payment Terms: LC, T/T, PayPal
Brand: ANJUE MEDICAL
Place of Origin: China
Display: Colored LCD, touch screen setting, big font display
Ventilation modes: SIPPV, SIMV, PCV, SPONT, MANUAL, CPAP

AJ-2202 High Quality Portable Medical Ventilator Portable Patient Ventilator Portable Ambulance Ventilator

Main technical parameters

DisplayColored LCD, touch screen setting, big font display, free changeable angles of views

Various ventilation modes SIPPV,SIMV,PCV,SPONT,MANUAL,CPAP

Ventilation parametersPEEP,SIGH,IRV,Inspiratory platform

Tidal Volume 50-1500 mL

Rate 1~99bpm

SIMVrate2~20bpm

I:E 4:1~1:8

Trigger sensitivity -1.0~1.0kPa

PEEP 0~25cmH2O

Pressure control3~50cmH2O

SIGH 1.5 times the inspiratory time (60 ~ 120 times adjustable)

Inspiratory platform 0~50%

Oxygen concentration48%~100%

Pressure limit5~60cmH2O

Parameters for ventilation monitoring tidal volume, ventilation volume, IPPV rate, total respiratory rate, peak pressure of airway, pressure - time waveform, flow rate - time waveform, inspiratory trigger pressure, Inspiratory platform, Oxygen concentration, battery capacity

Safety alarm system

Airway pressure alarm upper limit setting range0.3~6.0 kPa

lower limit setting range0.2~5.0 kPa

Per-minute ventilation volume alarm upper limit setting range 3.0~30L/min

lower limit setting range 1.0~10L/min

Vt alarm upper limit setting range 10~2000ml

lower limit setting range 0~1800 ml

Sustained high-pressure alarm it will give alarm when stress have consistently been higher than 2.5 kPa

Suffocation alarm it will give sound and light alarm if there is no tidal volume input for 15 seconds

Power failure alarm

PowerAC 220V±10% 50Hz±1Hz

UPS, Storage battery

Battery under-voltage alarm

Power loss alarm

Intubation off alarm