Business Type:
Manufacturer/Factory
Establishment:
2012
R&D Capacity:
ODM, OEM, Other
OEM/ODM Service
Sample Available

Hematology Analyzer, Medical Equipment, Analyzer manufacturer / supplier in China, offering Super Quality Auto Hematology Analyzer, Infant CPAP Ventilator, Mother  & Fetal Monitor  and so on.

1
    YRS
General Supplier
Supplier Homepage Furniture Outdoor Furniture Anesthesia Machine with Ventilator AJ-2105 Surgery Medical breathing assistant

Anesthesia Machine with Ventilator AJ-2105 Surgery Medical breathing assistant

Get Latest Price
Min. Order / Reference FOB Price
1 Unit US $7000.00/ Unit
Local Area: Anqing, Anhui, China
R&D Capacity: OEM, ODM, Other
Payment Terms: LC, T/T, PayPal
Brand: ANJUE
Place of Origin: China
Instrument classification: Class II
Power AC: 220V±10% 50Hz±1Hz (PS, Storage battery)

Anesthesia Machine with Ventilator AJ-2105

AJ-2105 Anesthesia Machine, with its ergonomic design, integrates a concise and smooth structure and high mobility. The microcomputer-controlled Ventilator adopts exported sensors and other components. Its scientific and proper design as well as rigorous production process is what demonstrates the inner quality of Anjue products: mechanical ventilation is made secure and visible by the folded respiratory bellows, and the security of air supply is guaranteed by pneumatic transmission and measurement system. With multiple modern technologies applied, the Anesthesia Machine is both outstanding in performance and versatile in functions.

The Ventilator adopts microcomputer control with high-definition 8.4” TFT color LCD display, integrating the functions of respiratory control and displaying the monitoring parameters. It is versatile in function and excellent in performance.

The high-precision Vaporizer is designed with a stable concentration output, automatic compensation for pressure, temperature and flow rate. And it is available to choose two kinds of gas from Halothane, Enflurane, Isoflurane, and Sevoflurane, ensuring a successful anesthesia operation.

High-definition five-tube flowmeter (oxygen, nitrous oxide, and air), designed with stoppers for oxygen and nitrous oxide to control the output oxygen concentration.

Reliable pneumatic transmission system.

Economic and proper configuration of technologies, optional as required by the Customer.

Modularized design and upgrading. Upgrading is available as required by the Customer.

TECHNICAL PARAMETERS

Part of Anesthesia Machine

Main body High-strength engineering plastic rack, light,

beautiful and corrosion resistant

Scope of applicationadult, child

Gas source O2: 0.27~0.55MPaN2O: 0.27~0.55MPa

Flow meter O2 : 0.1~1.0L/min1.1~10L/min

N2O: 0.1~1.0L/min1.1~10L/min

Air : 0.1~10L/min

O2, N2O linkage and N2O Stopper

When using nitrous oxide, oxygen concentration>25%;

When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.

Flow rate of rapid oxygen supply25~75L/min

Low oxygen pressure alarmThere will be sound alarm when the oxygen pressure < 0.2MPa

Vaporizer It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane,Enflurane, Isoflurane and Sevoflurane, two can be chosen for application as required by the customer. And vaporizers that are imported with original packaging are also available.

Respiratory circuitworking mode: all-close, semi-close, semi-open

APL: 0.5~7 kPa

Respiratory Bellowsbellows for adults, tidal volume range:0~1500 ml

bellows for children, tidal volume range:0~300 ml

anyone can be chosen for application as

required by the customer

Part of Ventilator

Display modeHigh-definition 8.4” TFT color LCD screen display

Ventilation mode IPPV, SIPPV, SIMV, PEEP, MANUAL,

SIGH, IMV, VCV, PCV

Ventilation parameters Tidal volume 20~1500 ml

Rate1~99 bpm

SIMV rate 2~20 bpm

I:E 4:1~1:8

Inspiratory trigger pressure-1.0~2.0 kPa

PEEP0~2.0 kPa

Pressure range 1.0~6.0 kPa

SIGH 1.5 times the inspiratory time

(70-150 adjustable)

Inspiratory platform 0~50%

Parameters for ventilation monitoringtidal volume, ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway, average pressure, pressure - time waveform, flow rate - time waveform,PEEP,inspiratory trigger pressure, Inspiratory platform, autonomous inspiratory rate, lung compliance, airway resistance, flow-volume loop, pressure-volume loop

Monitoring of oxygen concentration 21%~100%

Security Alarm System

Oxygen Concentration alarm upper limit setting range 21%~100%

low limit setting range10%~80%

Airway pressure alarmupper limit setting range 0.3~6.0 kPa

low limit setting range 0.2~5.0 kPa

Per-minute ventilation volume alarmupper limit setting range3.0~30L/min

low limit setting range1.0~10L/min

Sustained high-pressure alarmit will give alarm when stress have consistently

been higher than 2.5 kPa

Suffocation alarm it will give sound and light alarm if there

is no tidal volume input for 15 seconds

Power alarm

Power AC 220V±10% 50Hz±1Hz

(UPS, Storage battery)

(Mechanical Arms)

(End-expiratory CO2 monitoring)

*Items within brackets are optional.