100mA Medical Diagnostic X-ray Unit
X-ray GeneratorRotary anode, double focus, and bridge silicon rectification. The X-ray generator is an oil-immersed self-cooled unit.
Max rated capacity
Fluoroscopy: 75kVp, 3mA
Current 15mA 30mA 60mA 100mA
Voltage 90kVp 90kVp 90kVp 80kVp
Time 6.3s 6.3s 4.0s 3.2s
Power supplyVoltage: 180~240V; Frequency: 50Hz
Power: Not lower than 7kVA; Resistance: Not larger than 1Ω
Voltage adjustment range of X-ray tube45~90kVp, continuous adjustment
Range of the timer0.08~6.3s
Fluorescent screen size356 mm x 356 mm
Distance between the focus to the screenmax: 830 mm; Min: 630 mm
Tube head carriage movement rangevertical: 1100 mm; Horizontal: 260 mm
Beam limiterfield coverage at 100 cm FFD, 45x 45 cm (max); 0x 0 cm (min)
Specification of X-ray tubeModel XD55 rotation anode, double focus 1.8/4.3mm
Movement of the tabletop400mm
Field size of spot film device127mmx 178mm (1 pc); 203mmx 254mm (1 pc);
1/2 (127mmx 178mm) (2 pcs)
Transport dimension (Lx Wx H) (mm)2310x 1000x 1250
Weight (kg)Net: 418 / Gross: 564
X-ray GeneratorRotary anode, double focus, and bridge silicon rectification. The X-ray generator is an oil-immersed self-cooled unit.Max rated capacityFluoroscopy: 75kVp, 3mACurrent 15mA 30mA 60mA100mA
Voltage 90kVp90kVp90kVp80kVp
Time 6.3s6.3s4.0s3.2sPower supplyVoltage: 180~240V; Frequency: 50Hz
Power: not lower than 7kVA; Resistance: not larger than 1ΩVoltage adjustment range of X-ray tube45~90kVp, continuous adjustmentRange of the timer0.08~6.3sFluorescent screen size356 mm x 356 mmDistance between the focus to the screenmax: 830 mm; min: 630 mmTube head carriage movement rangevertical: 1100 mm; horizontal: 260 mmBeam limiterfield coverage at 100 cm FFD, 45x45 cm (max); 0x0 cm (min) Specification of X-ray tubeModel XD55 rotation anode, double focus 1.8/4.3mmMovement of the tabletop400mmField size of spot film device127mmx178mm (1 pc); 203mmx254mm (1 pc);
1/2 (127mmx178mm) (2 pcs) Transport dimension (LxWxH) (mm)2310x1000x1250Weight (kg)Net: 418 / Gross: 564