Veterinary Ultrasound Scanner, Bovine Ultrasound, Animal Ultrasound manufacturer / supplier in China, offering Bovine Pregnancy Instrument Handheld B Mode Veterinary Ultrasound Scanner, Portable 21 Inch HD LCD Screen Medical Monitor, Minimal Invasive Endoscope Medical Monitor and so on.
Min. Order / Reference FOB Price | |
---|---|
1 Piece | US $1800.00/ Piece |
Local Area: | Wuhan, Hubei, China |
---|---|
R&D Capacity: | OEM, ODM, Other |
Payment Terms: | LC, T/T, PayPal |
Type: | Anaesthesia Machine |
---|---|
Kind: | Gynaecology |
Certification: | CE, ISO13485 |
DP- 01Sta Factory price medical icu anesthesia machine anesthesia machine price
TECHNIQUES SPECIFICATIONS
Physical specifications
Screen
LED Display Screen
Suitable
Adult & Child
Mode
pneumatically powered and electrically controlled system
Working Mode
Closed; Semi-Closed; Semi-Open
Circuit
Breathing circuit integrated standards
Tube
2 Tubes Flowmeters: O2:0.1~10L/Min, N2O:0.1~10L/Min
Trolley
Fitted with 4 nos anti-static rubber castors; two of which are lockable for braking and easy maneuverability with foot operated brake provisions
Drawer unit
ONE drawer totally extractable
Technical Specifications
Gas requirement
Medical oxygen and nitrous oxide with a pressure ranging from O2: 0.32~0.6MPa ; NO2: 0.32 MPa to 0.6 MPa.
safety valve
<12.5 kPa
Respiratory Rate
4~40bpm
Flow meter
O2:0.1~10L/Min, N2O:0.1~10L/Min
oxygen concentration in the mixed gas N2O/O2
> 21%
Oxygen Flush
25~75 L/min
Maximum error of the pressure testing device
not exceed ± (4% of the full scale reading + 8% of the actual reading)
modes of ventilation
A/C,IPPV, SIPPV, MANUAL
I/E Ratio
1:1.5; 1:2.0;1:2.5; 1:3
Tidal Volume
0~1500ML
Controlled ventilation and assisted ventilation switching time
6s
Maximum safety pressure
≤ 12.5 kPa
Pressure limit range
0 ~ 6.0 kPa
Airway pressure alarm: Audible and visual and with yellow and red color indicating
Lower: 0.2kPa~5.0kPa; Upper: 0.3~6.0 kPa
±0.2 kPa
Tidal volume alarm: Audible and visual and with yellow and red color indicating
upper alarm: 50 to 2000ml,lower alarm: 0~1800ml
Power Supply Alarm
Ac/dc power supply are after failing to send out alarm immediately
Alarm Time: keep >120s
Operating conditions
Ambient temperature
10 ~ 40oC
Relative humidity
no higher than 80 %
Atmospheric pressure
860 hPa ~ 1060 hPa
Power requirement
100-120 Vac, 50/60 Hz;
Air source requirement
Medical oxygen and laughing gas with a rated pressure ranging from 0.3 to 0.5MPa.
Attention
the AC power supply used for the anesthesia machine must be well grounded.
Attention: the anesthesia machine used must be equipped with a carbon dioxide monitor complying with ISO 9918:1993, an oxygen monitor complying with ISO 7767:1997 and an expiratory gas volume monitor complying with 51.101.4.2 of Medical Electrical Equipment Part II: Special Requirements for Safety and Basic Performance of Anesthesia System.
Configuration
Vaporizer
Enflurane/ Isoflurane(Option: Halothane)
Storage
Ambient temperature
-15oC ~ +50oC
Relative humidity
no higher than 95 %
Atmospheric pressure
86 kPa ~ 106 kPa.