EC/TDS/Salinity/Temp Portable MeterDW-3010M
Features
Simultaneous displays of conductivity, salinity or TDS with temperaturemeasurements.Convenient one point calibration. Calibration data is stored in memory andis ready for use on power up.Selectable cell constants for true wide range readings.Adjustable TDS coefficient, reference temperature, temperature coefficient. Uses one conductivity cell for conductivity, salinity and TDS measurements.Accepts 2-wire conductivity cells with k=0.1, k=1.0 and K=10.0 cell constants for improved flexibility and accuracy.IP65 waterproof.Low battery indicator.
Specifications RangeConductivity0.00~20.00uS/cm; 20.0~200.0uS/cm; 200~2000uS/cm; 2.00~20.00mS/cm; 20.0~200.0mS/cmTDS0.000~2.000g/L; 0.00~20.00g/L; 0.0~200.0g/LSalinity0.0~ 80.0pptTemperature-10.0~120.0ºCResolutionConductivity0.01uS/cm; 0.1uS/cm; 1uS/cm; 0.01mS/cm; 0.1mS/cmTDS0.001g/L; 0.01g/L; 0.1g/LSalinity0.1pptTemperature0.1ºCAccuracyConductivity±0.5%FSTDS±0.5%FSSalinity±1%FSTemperature±0.5ºCReference TemperatureFrom 15.0 to 25.0ºC, default is 25.0ºC,Temperature CoefficientFrom 0.00 to 4.00%, default is 1.91%.TDS ConstantFrom 0.30 to 1.00, default is 0.65.Electrode Type / Cell ConstantTwo wire conductivity cell / 0.1, 1.0, 10, default is 1.0.ATCThermistor, 10K ohms, at 25ºC.Power9 Volt battery.Operating Ambient ConditionsFrom 0~50ºC, at 90% RH.Weight / Dimensions (W x D x H)260 grams (Including batteries) / 70mm x 186mm x 37mmAccessories109ATwo-wire conductivity cell with 4 PIN connector, K=0.1 (0.0~200.0uS/cm).106ATwo-wire conductivity cell with 4 PIN connector, K=1.0 (0.02~20.00mS/cm).107ATwo-wire conductivity cell with 4 PIN connector, K=10.0 (1.0~200.0mS/cm).