Business Type:
Manufacturer/Factory
Business Range:
Ultrasound Detecting Equipment, Ultrasound Series Vascular Doppler Detectors
Establishment:
2001
R&D Capacity:
ODM, OEM, Other
OEM/ODM Service
Sample Available

Ultrasound, Ultrasound Scanner, Ultrasound System manufacturer / supplier in China, offering Ultrasound Scanner (BEU-8360), Needle Destroyer Syringe Destroyer Disposable Syringe Destroyer Bd-310 with Patent Since 2003, Blood Warmer, Fluid Infusion Warmer, Temperature Precise Control and so on.

1
    YRS
General Supplier

China Fetal Monitor

Get Latest Price
Min. Order / Reference FOB Price
1 Production Contact Supplier
Local Area: Shenzhen, Guangdong, China
R&D Capacity: OEM, ODM, Other
Payment Terms: LC, T/T, PayPal
Classification: Physiological Functions of Diagnosis and Monitoring Equipment
Type: Patient/Fetal/Maternal Monitor
Feature: Can Be Connected to Computer
Features:
5.7" LCD display can display main parameters for US and TOCO Data, ECG-MHR, TEMP, RR, SpO2, NIBP and monitoring curve;
FHR collection signal for quality indication, alarm indication for abnormal FHR, Fetal movement indication;
Coding keyboard adjusts all parameter setting of unit, memory and replay the data in 24 hours(for option);
Record FHR, UC, FM mark, necessary character interpretation and monitoring time synchronously with 832 thermal linear array printer;
Curve printing speed, printing line of thin or thick, signal collection rate can be adjust for the requirements of user, the printing speed can be setup to 1, 2, 3cm/min;
Adjust the upper and lower alarming limit, real time, date and other parameter, the set-up data can be saved automatically after powered off;
The monitoring data can be conveyed to the exterior computer by RS-232.
Specifications:
FHR
Probe: 9 crystals, wide wave beam
Working mode: Pulsed Doppler
Ultrasonic frequency: 1.0MHz± 10%
Ultrasonic average intensity: ≤ 10mW/cm2
Signal processing: Signal processing units for special digital signal
FHR range: 50- 240bpm
Alarm range: Up: 55-240bpm Low: 50-235bpm
Printing resolution: 27bpm/cm
The highest frequency output: ≤ 5W
Overall sensitivity: ≥ 90dB
TOCO
Display range: 0-99units ((nonlinear error ≤ ± 10%)
Printing resolution: 40units/cm
Display: Display monitoring status for US and TOCO curve, fetal movement mark, doctor mark, time, volume, and NIBP, SpO2, ECG and up to and freeze the data
Battery: Ni-MH 16.8V/1800mA rechargeable batteries group
Time of continue working in static condition: ≥ 1 hours
Power supply: AC 100--240V 50Hz/60Hz DC 24V± 1V
Consume power: ≤ 50W
ECG and HR
Measurement Range: 30--350bpm
Measurement accuracy: 1bpm or ± 1%, get the bigger data
Cardiac electric noise level: ≤ 30μ V
Cardiac electric frequency response:
Filtering mode: 1--25MHz
Non-filtering mode: 0.05--100MHz
Cardiac electric Input loop current: ≤ 0.1μ A
NIBP
Measurement range:
E) SBP: (5.3--36.0) kPa or (40--270) mmHg
F) DBP: (2.7--26.7) kPa or (20--200) mmHg
G) MAP: (4.0--29.3) kPa or (30--220) mmHg
H) measurement accuracy: ≤ ± 0.4 kPa (3mmHg) or ± 2%, get the highest data
RR
Measurement range: 0--120bpm
Measurement accuracy: ± 5% get the highest data
TEMP
Channel for TEMP: 1
TEMP probe: Surface probe 1
Measurement range: 25--45º C
Measurement accuracy: ± 0.2º C
Response time: ≤ 30s
SpO2
Measurement range: 0--100%
Measurement accuracy: 70--100% (± 5%)
50--69% (± 3%)
PR
Measurement range: 30--250bpm
Measurement accuracy: 1bpm or ± 2%, get the highest data
Working temperature: 10--40º C
Humidity: ≤ 80%
Atmospheric pressure: 86--106kPa
Size: 252cmx224cmx190(mm)