Fully Automated Elisa Microplate Open System ELISA Processor
Total 25 racks, arbitrary rack can be defined as sample or reagent on the software.
4 units 96 well microplates (pipetting,incubator,reading).
Model
BIObase4000
Sample Pipetting Unit
Sample Rack
24 sample tubes/rack, at most 23 racks
Primary Sample Tubes
11.5-13mm
Pipetting
8ul~1000ul, (10ul) CV<5%
Time To Dispense
8 minutes adding 96 samples
Dispensing System
1 aspirating-dispensing probe with X-Y-Z movement;dilution system with a syringe of 1000ul, continuous flow washing station.
Reagent Pipetting Unit
Reagent Racks
2 reagent tanks(60ml)/ rack, at most 23 racks
QC Racks
24 positions/QC rack, at most 23 racks
Pipetting
8ul~1000ul,1ul stepping
Pipetting Precision
(100ul )CV≤0.5%
Time To Dispense
2 minutes(typical)^100ul of 96 reagents to plate from reagent bottle
Washing Unit
Programmable
Yes(options & parameters)
Wash Containers
3 wash buffers at 2.0L, 1 wash buffers at 5.0L,with level-sensing
Waste Containers
25L with waste full sensor
Washing System
two 8 nozzle manifolds, one to dispense and the other one to aspirate
Washing Residual
<2μl
Reading unit
Reading System
8 independent photometric channels with mono and bichromatic reading capability
Photometric Range
0.000 ~3.000 OD
Spectral Range
400-700nm
Optical Filters
2 precision filters (450and 630nm); 2 more on request (4 filters max, 405 nm, 450 nm, 492 nm and 630 nm)
Reading Inaccuracy
inaccuracy±1% from 0.000 to 1.500 OD, ±2% from 1.500 to 3.000 OD
Incubation unit
Incubators
4 independent units
Programmable
Yes
Temperature Range
RT to 45 °C
Temperature Accuracy
±0.5ºC
Temperature Staility
±0.5ºC
Software unit
Software System
Windows 7 /8/10 or above, LIS system available
Function Application
Work list set up, patient reports and result filing; calibration curve ,storage for each carried out test
Data Reduction
Cut-off (qualitative)
Physical Specification
Power Supply
AC220V+ 10%,50/60Hz,450W
Instrument Dimension (L*W*H)
1235*710*880mm
Package Dimension (L*W*H)
1370*850*1439mm
Total Weight
220kg