High-End Export Medical Hospital Anesthesia Machine (THR-MJ-S6100D)
Adopts the electric control gas driven mode.With a variety of alarm, presentation and protection functionUse 7 "color LCD display shows ventilation parameters, alarm information, and the graphical display.For adults and childrenO2, N2O flowmeter has the linkage,ensure O2 concentration≥25%;With O2 drive insufficient pressure alarm and N2O truncation protectionAir source O2,N2OAir pressure280 kPa ~ 600 kPa.Power supply 100-240V~Frequency 50 /60HZInput power 40 VA
Intermittent positive-pressure ventilation (IPPV)Assist / Control (A/C)Synchronized Intermittent Mandatory Ventilation (SIMV)Main technical parametersFlow meterO20.1~10 L/min
N2O0.1~10 L/min
O2 and N2O have internal linkage devices, ensure O2 concentration 25% Oxygen flush 35 L/min~ 75 L/minTidal volume(Vt) 0 mL ~ 1500 mLMV For adult≥18 L/min Frequence(Freq) 4 bpm ~ 40 bpm(SIMV mode:1 bpm~ 20 bpm)I/E 2:1 ~ 1:6Press triggering sensitivity(PTr)-20 cmH2O ~ 0 cmH2O(based on PEEP)Sigh Ventilation (Sign)0(off), 1 ~ 5 (/100)Pressure Limit20 cmH2O ~ 100 cmH2O
Frequence(Freq) 0 /min ~ 100 /minTidal volume(Vt) 0 mL ~ 2000 mLMV0 L/min ~ 100 L/minairway pressure0 cmH2O ~ 100 cmH2O
P-T (Airway Pressure -- Time Waveform)
The AC power outage alarmNo tidal volumeHigh (low) ventilation alarmHigh (low)Airway pressure alarmPressure limitContinuous pressure alarm The maximum limited pressure <12.5 kPaWith O2 drive insufficient pressure alarm and N2O truncation protectionO2, N2O flowmeter has the linkage to ensure O2 concentration≥25%.When O2 closed ,N2O closes at the same time.
Vaporizer
Anesthesia gasAdjustable scope % (volume percentage)Halothane0 ~ 5Enflurane0 ~ 5Isoflurane0 ~ 5Sevoflurane0 ~ 8