High Frequency Electrosurgical Generator
Model: S900C
Working modes: Cut and coagulation.
itemtypeS900Cdetailworking modespure cut or cut5~300W(may provide type 5~400W)blend 1-blend 2-blend 3-desiccate-fulgurate or coag.5~150Wspray-bipolar-range of use
May use to various surgical operations of cutting or/and coagulation to organism tissue.disposation of accessoriesactive electrode5footswitch1neutral electrode1bipolar electrode-disposable electrodepick out and buytransport datapackage dimensions45x26.5x47cmgross weight11kgtechnology datavoltage of mains220V ±22V,50Hz/60Hzworking frequency512kHz ±5Hzvariety of equipmentclass 1,type CF. ordinary equipmentworking fashionLoading intermittently, functioning continuously.leskage currents to groundless than 0.5mAleskage currents through patientless than 0.01mA