Business Type:
Trading Company
Business Range:
Mask Tester, Particle Filtration Efficiency Tester, Meltblown Particle Filtration
Establishment:
2017
R&D Capacity:
OwnBrand(Scince)
OEM/ODM Service
Sample Available

Tester, Mask Tester, Mask Breathing Resistance Tester manufacturer / supplier in China, offering Mask Breathing Resistance Test Equipment for Exhalation Resistance and Inhalation Resistance Testing, Portable Melt-Blown Fabric and Medical Face Mask Test Instrument for Filtration Efficiency and Resistance Testing, Automated Face Mask and Meltblown Filter Tester for Particle Filtration Efficiency and Resistance and so on.

1
    YRS
General Supplier
Supplier Homepage Instruments & Meters Test Instrument Test Machine China Melt Blown Tester for Meltblown Nonwoven Fabric Testing

China Melt Blown Tester for Meltblown Nonwoven Fabric Testing

Get Latest Price
Min. Order / Reference FOB Price
1 Set US $9500.00/ Set
Local Area: Qingdao, Shandong, China
R&D Capacity: OEM, ODM, Other
Payment Terms: LC, T/T, PayPal
Test Flow: 32 L/Min
Flow Meter Accuracy: ±3%
Pressure Gauge Accuracy: 1PA
Melt Blown Tester for Meltblown Nonwoven Fabric Testing SC-MBT-2032
Application
This instrument is used to quickly detect the filtration efficiency and resistance of melt-blown, flat masks at 32L/min flow rate.
Technical Parameters

Technical Data SheetItemParameterTest flow32L/minTest area100 cm2Test particle size0.3μmFlow meter accuracy±3%Pressure gauge accuracy1PaAerosolPaticulates in the airDetectorLaser particle counterPower supplyAC220V, 50 or 60 Hz, can be customizeDimension300mm*250mm*360mm


Product Pictures

Features
1. Ambient air is used as the detection medium, calibration filter paper is provided, and the filtration efficiency of 0.3 micron NaCl particles under standard conditions is obtained by adjusting the balance coefficient.

2. Used for relative comparison, not absolute judgment. As a non-standard test equipment, environmental impact may affect the test results. Therefore, the best use of this portable instrument is for the relative comparison of melt-blown materials, rather than the absolute value of the test results as the basis for judgment.

3. At the same time, in order to achieve the accuracy of the absolute value of the test result as much as possible, the instrument is designed with balanced correction, which is used to correct the absolute value of the test result in order to be closer to the standard instrument.
4. The resistance (Pa) of the test result is not affected by the environment, and the test result can be compared with the standard instrument.
5. This instrument can be used as a resistance tester for semi-finished products of melt-blown plants, and can also be used as a product uniformity testing device in product control of melt-blown plants. The cost is low and the operation is easy.