Model No | GTO-080 | GTO-150 | GTO-225 | GTO-408 | GTO-800 | GTO-1000 |
Working room(L) | 80 | 150 | 225 | 408 | 800 | 1000 |
Inner chamber size(mm)W*H*D | 400*500*400 | 500*600*500 | 500*750*600 | 600*850*800 | 1000*1000*800 | 1000*1000*1000 |
Exteriorchamber size(mm)W*H*D | 900*900*1500 | 950*1500*1050 | 950*1650*1150 | 1050*1750*1350 | 1450*1900*1350 | 1450*1900*1550 |
Packaging volume(CBM) | 2 | 3 | 3.5 | 4.5 | 5.5 | 6 |
G.W.(KGs) | 300 | 320 | 350 | 400 | 600 | 700 |
Heating&CoolingRate | 3°C/min,5°C/min,8°C/min,10°C/min,15°C/minlinearaverageornonlinearaverage |
Performance | Temperature range | -160ºC,-150ºC,-120ºC,-100ºC,-80ºC,-70ºC,-60ºC,-40ºC,-20ºC, 0ºC~+150ºC,200ºC,250ºC,300ºC,400ºC,500ºC |
Temp.fluctuation | ±0.5°C |
Temp.Uniformity | ±1.5°C,No-load operation,After stead state 30 min. |
Temp.resolution | 0.01°C |
Heating rate | ≥3°C/min |
Cooling rate | ≥1Industrial D Printer °C/min |
Material | Inner chamber material | SUS#304 stainless steel plate |
Exterior chamber material | Stainless steel plate+ powder coated |
Insulation Material | PU & Fiberglass wool |
System | Air circulation system | Cooling fan |
Fan | Sirocco fan |
Heating System | SUS#304 stainless steel high-speed heater |
Air flow | Forced Air Circulation(It enters at bottom and leaves at top) |
Refrigeration system | imported compressor,Tecumseh compressor(or Bizer Compressor),finned type evaporator,air(Water)-cooling condenser |
Refrigerating fluid | R23/ R404A USA Honeywell. |
Condensation | Air(Water)-cooling condenser |
Controlling system | Digital electronic indicators+S.S.R.With PID automatic calculation capability |
Operation interface | Grande Expertise in Temperature & Humidity Controller,Chinese-English Shift. |
Controller | Programmable capability | Save 120 profiles with up to 1200 steps each |
Setting range | Temperature:-100ºC+300ºC |
Reading accuracy | Temperature:0.01ºC |
Input | PT100 or T Sensor |
Control | PID control |
Communication interface | Equipped with standard communication interface devices USB,RS-232 and RS-485,enable the test chamber to be connected with personal computer(PC),to achieve multi-machine control and management at the same time.Standard: USB external memory port.Optional: RS-232, RS-485, GP-IB, Ethernet |
Print function | Japan Yokogawa Temperature Recorder(Optional accessories) |
Auxiliary | Limit Alarm,Self Diagnosis,Alarm Display(Failure Cause),Timing device(Automatic Switch) |
Accessories | Multi-layer vacuum glass observation window,Cable port(50mm),Controlling status indicator lamp,Chamber light,Specimen loading shelf(2pcs,position adjustable),Guaze 5pcs,Operation manual 1 set. |
Safety protection device | Over-heat protection circuit breaker,Compressor overload protection,Control system overload protection,Humidifying system overload protection,Overload indicator lamp. |
Power supply | AC 1Ψ 110V;AC 1Ψ 220V;3Ψ380V 60/50Hz |
Customized service | Welcome to Non-standard,Specail requirements,OEM/ODM orders. |
The technical information will be subjected to change without notice |