Computerized High-Frequency First-Aid Ventilator AJ-2204
This is a dedicated system device for respiration salvaging and revival, The main frame choose the principle of high-frequency ventilation, it is widely used for clinical salvaging and revival also is a dedicated device used for respiration salvaging on first-aid car.
Features
.An opening system is designed to avoid leakage of airway, When ventilating, the intubation may without casing bag. Little tube may be used for ventilation when resistance of expiration is small, so that mucosa won't be hurt easily.
With opening airway, suction and ventilation can be done at the same time, so good effectiveness of oxygen combination can be shown without inspiration and expiration synchronizing.
As ventilating, lower airway pressure is good to increase blood discharge and decrease encephal pressure.
Suction device, it driving negative pressure is adjustable.
Accumulator jar, it can be used continuously for not less than four hours after being charging.
The system disposes completely with laryngoscope, inlet intubation, catheter of tracheotomy, nasal prongs tube, air tube and so on.
Main technical parameters
Output power: ≤65VA Power supply: AC220V+22V 50HZ+1HZ
Frequency:10~150 breaths/min
I/E ratio:1:1, 1:1.5~1:3 and inverse ratio
SIGH:6 breaths/min
CPAP: Not less than 1.5KPa
Suction negative pressure:-0.01~0MPa
Driving Pressure:0~0.15MPaRelative humidity : ≤80%Temp.: 5-40%
Pressure Regulator Limit:0.30MPa
Constant volume ventilation: T=10min,I/E=1:1.5
Adults: f(100):f(20)=7:3 tidal volume=550ml/timeDriving pressure=0.06MPa
Children: f(100):f(20)=7:3 tidal volume=350ml/timeDriving pressure=0.04MPa
First-aid: 1) adult T=10min, F(100)=7min, F(20)=3min, Tidal=550ml/time, Cycle
2) Child T=10min, F(150)=7min, F(30)=3min, Tidal=350ml/time, Cycle
Storage battery:4Ah,Automatic charge when 12V
Most jetting volume: ≥1400ML/time
Oxygen concentration: 35-90% adjustable
Humidifying ratio: 0-15ml/hour
Airway pressure display: 0-3Kpa
Alarm: Alarm when battery pressure is lower than 11V;Alarm when the oxygen pressure is lower than 0.01MPa;Alarm with sound and light when no gas
Ventilation Mode: HFJV, CPAP, HFJV+SIGH(Sigh), Constant volume ventilation.