Supplier Homepage Furniture Outdoor Furniture First Aid Ventilator AJ-2204 HFJV CPAP HFJV+SIGH

First Aid Ventilator AJ-2204 HFJV CPAP HFJV+SIGH

Get Latest Price
Min. Order / Reference FOB Price
1 Unit US $4000.00/ Unit
Local Area: Anqing, Anhui, China
R&D Capacity: OEM, ODM, Other
Payment Terms: LC, T/T, PayPal
Brand: ANJUE
Place of Origin: Anhui, China
Frequency: 10~150 breaths/min
Driving Pressure: 0~0.15MPa

Computerized High-Frequency First-Aid Ventilator AJ-2204

This is a dedicated system device for respiration salvaging and revival, The main frame choose the principle of high-frequency ventilation, it is widely used for clinical salvaging and revival also is a dedicated device used for respiration salvaging on first-aid car.

Features

.An opening system is designed to avoid leakage of airway, When ventilating, the intubation may without casing bag. Little tube may be used for ventilation when resistance of expiration is small, so that mucosa won't be hurt easily.

With opening airway, suction and ventilation can be done at the same time, so good effectiveness of oxygen combination can be shown without inspiration and expiration synchronizing.

As ventilating, lower airway pressure is good to increase blood discharge and decrease encephal pressure.

Suction device, it driving negative pressure is adjustable.

Accumulator jar, it can be used continuously for not less than four hours after being charging.

The system disposes completely with laryngoscope, inlet intubation, catheter of tracheotomy, nasal prongs tube, air tube and so on.

Main technical parameters

Output power: ≤65VA Power supply: AC220V+22V 50HZ+1HZ

Frequency:10~150 breaths/min

I/E ratio:1:1, 1:1.5~1:3 and inverse ratio

SIGH:6 breaths/min

CPAP: Not less than 1.5KPa

Suction negative pressure:-0.01~0MPa

Driving Pressure:0~0.15MPaRelative humidity : ≤80%Temp.: 5-40%

Pressure Regulator Limit:0.30MPa

Constant volume ventilation: T=10min,I/E=1:1.5

Adults: f(100):f(20)=7:3 tidal volume=550ml/timeDriving pressure=0.06MPa

Children: f(100):f(20)=7:3 tidal volume=350ml/timeDriving pressure=0.04MPa

First-aid: 1) adult T=10min, F(100)=7min, F(20)=3min, Tidal=550ml/time, Cycle

2) Child T=10min, F(150)=7min, F(30)=3min, Tidal=350ml/time, Cycle

Storage battery:4Ah,Automatic charge when 12V

Most jetting volume: ≥1400ML/time

Oxygen concentration: 35-90% adjustable

Humidifying ratio: 0-15ml/hour

Airway pressure display: 0-3Kpa

Alarm: Alarm when battery pressure is lower than 11V;Alarm when the oxygen pressure is lower than 0.01MPa;Alarm with sound and light when no gas

Ventilation Mode: HFJV, CPAP, HFJV+SIGH(Sigh), Constant volume ventilation.