Microplate reader Specification:
Large LCD Display, size is 240x 128
Input: Touch screen
Detection method: Absorbance
Microplate types: 96-well plate(U, V or flat-bottom)
Measuring system: 8-Channel optical ststem
Reading speed: 12s single wavelength, 30s double wavelength
Light source: Tungsten Halogen, 12V/20W, >2000hours
405nm, 450nm: 492nm, 630nm, 4free positions for additional filters
onboard software: 42user-programmable protocols
Reading range: 0.001-3.500Abs
Measurement range: 0.001-2.500Abs
Photometric accuracy: <1% or ± 0.001Abs
Wavelength accuracy: <2nm
Repeatability: <1%
Stability: <0.005Abs/10min
Linearity: <2.0%
Shaking: Linear shaking; Low, medium, high
Power supply: AC220V/110± 10%, 50/60Hz
Operating environment: Temperature: 10° C ~35° C
Humidity: 15~85%
8-Channel optical measuring system
Optional single/double wavelength analysis
Variable speeds linear shaking
Cutoff and call criteria
Large storage for 2500 results
DisplayLargeLCD240x128InputTouchscreenDetectionmethodAbsorbanceMicroplatetypes96-wellplate(U,Vorflat-bottom)Measuringsystem8-ChannelopticalststemReadingspeed12ssinglewavelength,30sdoublewavelengthLightsourceTungstenHalogen,12V/20W,>2000hours405nm,450nm492nm,630nm,4freepositionsforadditionalfiltersOnboardsoftware42user-programmableprotocolsReadingrange0.001-3.500AbsMeasurementrange0.001-2.500AbsPhotometricaccuracy<1%or±0.001AbsWavelengthaccuracy<2nmRepeatability<1%Stability<0.005Abs/10minLinearity<2.0%ShakingLinearshaking;Low,medium,highPowersupplyAC220V/110±10%,50/60HzOperatingenvironmentTemperature10°C~35°CHumidity15~85%