Business Type:
Manufacturer/Factory,Trading Company
Business Range:
Medical Equipment, Ultrasound, X-Ray, MRI, C-Arm
Establishment:
2018
R&D Capacity:
OEM, ODM, Others
Terms of Payment:
LC, T/T, D/P, Paypal, Western Union
Main Markets:
Africa, South America, Eastern Asia
OEM/ODM Service
Sample Available

Our main business is medical products, Focus on to provide our clients with quality products and services, the company since its inception, with the vast number of customers to establish a good busine...

1
    YRS
General Supplier

ICU Ventilator 3000A PLUS

Get Latest Price
Min. Order / Reference FOB Price
1 Set US $140000.00/ Set
Local Area: Xian, Shaanxi, China
R&D Capacity: OEM, ODM, Other
Payment Terms: LC, T/T, D/P, Paypal, Western Union
Brand: Kindsway
Model Number: 3000A PLUS
Place of Origin: Jiangsu, China
Instrument classification: Class III

ICU Ventilator 3000A PLUS

Product Description

This ICU ventilator is akind of pneumatic driven microcomputer controlled ventilator.

It is mainly used in VCV, A/C, SIMV, PSV, Spont., MMV, PCV, PEEP, SIGH and other functions. The main components of the ventilator are imported of high quality products, which greatly improves the reliability and stability of the ventilator. The ventilator is reasonable in structure, convenient in operation and reliable in performance, and is suitable for respiratory rescue and treatment in emergency and ICU Wards.

Main performance and Technical Index

Ventilation mode: VCV, A/C, SIMV, PSV, Spont., MMV, PCV, SIGH, PEEP.

Air circuit Performance:

Maximum safety pressure of air circuit: no more than 6kpa.

Inspiratory oxygen concentration: 21-100%.

Ventilation Performance:

Respiratory rate (f): 4-99 times / min,

Inspiratory time (Ti): 0.2-6 s,

Breath holding time (Tp): 0-2s,

Triggering pressure (Pt): inspiratorytriggering pressure is - 2kpa-0kpa,

Tidal volume (Vt): 0-2000ml;

Maximum minute ventilation volume: not less than 18 L / min.

PSV:0~5 kPa,

Peep Pressure: 0-3kpa,

Inspiratory flow rate: 0-80 L / min.

SIMV frequency: 6-60 times / min,

Inspiratory pressure: 0-6kpa,

Alarm Performance:

Pressure High Limit (Ph) alarm: 2~ 6kpa, audible and visual alarm.

Pressure Low Limit (Pl) alarm: 0~2KPa, audible and visual alarm.

Minute ventilation High Limit (Mvh) alarm: audible and visual alarm when ventilation exceeds the set value.

Minute ventilator Low limit (Mvl) alarm: audible and visual alarm when ventilation is lower than the set value.

Power off alarm: Power off alarm immediately, and the alarm duration shall not be less than 120s.

Mute: no more than 2 minutes.

Monitoring Performance:

10.4 inch color LCD

Tidal volume (VT): 0 ~ 2000ml

Measured total frequency (Ftot), Spont. frequency (Fspont)

Minute ventilation (Mv).

I: E ratio.

Airway pressure (PAW).

Peak pressure (Ppeak)

Peep pressure: 0.1 ~ 2KPa

Pressure-Time and Flow-Time waveform