Real-time infused volume / bolus rate / bolus volume / KVO rate, automatic power switching, mute key, purge, bolus, system memory, key locker, change flow rate without stopping the pumpFeatures:
• Compact in design, light in weight and small in size.
• Compatible with universal IV set.
• Low motor driving noise.
• Ultrasonic bubble sensor.
• Easy to set VTBI (volume to be infused) by [INCR] or [DECR] key on the front panel.
• Accurate setting of flow rate for patients.
• Flow rate accuracy with the equipped peristaltic finger system.
• The infused volume can be cleared by pressing [CLEAR] key without switching off the power.
• Audio-visual alarms for added safety.
• The reminder alarm sounds repeatedly if no action is taken within 2 minutes after the alarm was switched off.
• The flow rate can be set in 0.1ml/h increments.
• After delivering the VTBI, the pump continues running with keep vein open (KVO rate) mode.
• When door is open, the tube is automatically locked by tube clamp.
• The rechargeable built-in battery allows the pump to be transported with the patient without ceasing normal pump
operation.
SpecificationsModelMVP NPumping MechanismCurvilinear peristalticIV SetCompatible with IV sets of any standardFlow Rate0.1-1500 ml/h (in 0.1 ml/h increments)Purge, Bolus100-1500 ml/h (in 1 ml/h increments)
Purge when pump stops, bolus when pump startsBolus volume1-20 ml (in 1 ml increments)Accuracy±3%VTBI1-9999 mlInfusion Modeml/h, drop/min, time-basedKVO Rate0.1-5 ml/h (in 0.1 ml/h increments)AlarmsOcclusion, air-in-line, door open, end program, low battery, end battery,
AC power off, motor malfunction, system malfunction, reminder alarmAdditional FeaturesReal-time infused volume / bolus rate / bolus volume / KVO rate,
automatic power switching, mute key, purge, bolus, system memory,
key locker, change flow rate without stopping the pumpOcclusion SensitivityHigh, medium, lowAir-in-line DetectionUltrasonic detectorWireless managementOptionalPower Supply, AC110/230 V (optional), 50-60 Hz, 20 VABattery9.6±1.6 V, rechargeableBattery Life5 hours at 30 ml/hWorking Temperature10-40ºCRelative Humidity30-75%Atmospheric Pressure700-1060 hpaSize174*126*215 mmWeight2.5 kgSafety ClassificationClass I, type CF
1) LCD display
2) Keys
3) Flow LED: Flash when infusion.
4) DIP switch: IV set selection.
5) Bubble sensor: Detects air bubble in the IV tubing.
6) Tension plate: Press on the tubing to generate pressure.
7) Fingers: Peristaltic press on the tubing to drive down the solution.
8) Door lever: Pull up this lever and close the door. Just push the door to lock it.
9) Tubing guider: Guide the IV tubing.
10) Pressure sensor: Detects the occlusion of the IV tubing.
11) Tubing clamp: Automatically clamps the IV tubing when the door is open.
12) Pump door.