Medical Freeze Dryer, Food Freeze Dryer, Fruit Freeze Dryer manufacturer / supplier in China, offering Bench-Top Mini Small Laboratory Vacuum Freeze Dryer (Lyophilizer) Drying Machine, Ftir Fourier Transform Infrared Spectrometer (High performance), Automatic Fusion Machine (for X-ray fluorescence spectroscopy) and so on.

1
    YRS
General Supplier
Supplier Homepage Instruments & Meters Test Instrument Other Test Instrument PT-D92-3536A Automatic Cleveland Open-Cup Flash Point Tester ASTM D 92

PT-D92-3536A Automatic Cleveland Open-Cup Flash Point Tester ASTM D 92

Get Latest Price
Min. Order / Reference FOB Price
1 Piece Contact Supplier
Local Area: Xian, Shaanxi, China
R&D Capacity: OEM, ODM, Other
Payment Terms: LC, T/T
Power Supply: AC220(-10%~+5%)V, 50Hz
Flash Point Determination: Range: Ambient to 400ºC
Heating Rate: Correspond with GB/T3536 and ASTM D92
Summary
This instrument is designed and made as per and . It is applicable to all petroleum products with flash points above 79ºC and below 400ºC except fuel oils.I. Main technical features
1. This instrument adopts LCD screen to display and full English man-machine interface. It can preset expected flash point temperature, sample mark number, barometric pressure and other parameters.It has menu prompt and input guide.
2. It adopts simulation tracking to display the function curve of temperature rising and test time. It has the functions of English mis-operation prompt,test date,test time and other parameters.
3. Automatically calibrate the effect of atmospheric pressure to test and calculate the corrected value.
4. Differential coefficient detection.Automatically correct the system deviation.
5. Automatically open the lid,detect the flash point and print test data.The test arm automatically rise up and lower down.
6. Reasonable design, safe and convenient to operate.II. Main technical specifications
1. Power supply:AC220(-10%~+5%)V, 50Hz
2. Flash point determination:Range:Ambient to 400ºC
Repeatability:≤8ºC
Reproducibility:≤17ºC
Accuracy:0.1ºC
3. Heating rate:Correspond with GB/T3536 and ASTM D92
4. Ignition mode:Electric ignition.
Gas flame diameter is 3.2mm~4.8mm
5. Ambient temperature:(10~40)ºC
6. Relative humidity:≤80%
7. Maximum power consumption:400W
8. Dimension:410mmx360mmx310mm
III.Standard AccessoriesNo.ItemUnitQtyRemarks1Oil cupPiece12Power supply wire(250V 6A)Piece13Printer paperRoll14IgniterPiece1