Erythrocyte Sedimentation Rate, ESR, Automatic manufacturer / supplier in China, offering Automatic Erythrocyte Sedimentation Rate (ESR-2020) Analyzer, Bio&Media Creatinine (CRE) Test Kit/Ivd Reagents/Hospital/Lab/Advanced Renal Disease/High-Efficiency, Bio&Media Urea (UA) Test Kit/Ivd Reagent/Uao-Pod Method/Hospital/Lab/High Stability/High-Efficiency and so on.
Min. Order / Reference FOB Price | |
---|---|
1 box | US $672.00/ box |
Local Area: | Jinan, Shandong, China |
---|---|
R&D Capacity: | OEM, ODM, Other |
Payment Terms: | LC, T/T |
Type: | Hcy Assay Kit |
---|---|
Quality Guarantee Period: | 12 Months |
Application: | Hospital, Lab, School |
1. Absorbance of blank
A (R1 reagent) ≥ 1.202. Sensitivity
The difference of absorbance (ΔA) ≥ 0.01 while testing the specific Hcy sample of concentration of 10.00 μmol/L.3. Linear range
(1) Linear correlation coefficient ∣r∣≥0.995
(2) Linear range of Hcy level: [3.00, 70.00] μmol/L
Linear rangeLinear deviation[3.00, 20.00]≤ ± 2.00 μmol/L[20.01, 70.00]≤ ± 10.00%
4. Precision
(1) Intra CV%
Concentration of Hcy (μmol/L) Intra CV%10.00 ± 2.00 ≤ 5.0%20.00 ± 4.00≤ 3.0%
(2) Inter-assay relative range is within 10.0% while testing the sample of 10.00 ± 2.00 μmol/L.
5. Accuracy