BNT-2000B Infant Radiant Warmer

STANDARD COMPOSITION:

Temperature Control Mode: pre-heating control, servo control & manual control;

Instant Parameter Display: set temp, skin temp, heat power rate;

Built-in APGAR Timer and >37ºC preset function

Audible/Visual Alarms: temp deviation, sensor failure, power failure, temp override;

Tilting Bed with Rotary Radiant Head;

X-ray Cassette Tray under infant bed, examination light of high illuminance;

Use anti-blast quartz infrared tube as radiant heating source;

With two LED phototherapy unit;

With two 7L oxygen cylinders, with pressure meter and flow meter;

With Suction, with neonate resuscitation.

TECHNICAL SPECIFICATIONS:

Temperature Control Range: 25ºC-37ºC,

Temperature override mode to: 38ºC

Temperature sensor accuracy: ≤0.3ºC

Mattress surface temperature uniformity: ≤2.0ºC

Temperature rise time: ≤45mins

Heat output indicate: 0-100% in 10%

Radiant box horizon angle adjust range: 0-90°adjustable in step-less

Bed tilt adjustment: continues adjustable in ±10°

Mattress Tray Size: 1700px*1350px

Mattress tray Max. load: 10kg

Transfusion pole Max. load: 2kg

Shelf Max. load:2kg

ALARM

Sensor failure alarm

Over temperature alarm

Deviation alarm: ±1ºC

Power failure alarm

APGAR time: alarm with audible when run between 50 seconds to 1 minute, 4 minutes 50 seconds to 5 minutes,

9 minutes 50 seconds to 10 minutes,

ENVIRonMENT TEMPERATURE

(It is a suggestion that please do not use outside of the allow environment.)

Operate temperature range: 18ºC~30ºC

Store temperature range: -40ºC~+55ºC

ENVIRonMENT HUMIDITY

Operate humidity range: 30%RH~75%RH

Store humidity range: ≤93%RH

ATMOSPHERIC PRESSURE

Transport and Store atmospheric pressure range: 500hPa~1060hPa

Operate atmospheric pressure range: 700hPa~1060hPa

STREAM VELOCITY

Ambit environment stream velocity: ≤0.3m/s